Zugriffsnummer 36098
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel Correction of shaker flatness deviations in very low frequency primary accelerometer calibration
Autor(in); Institution
Bruns, Thomas; 1.7, Akustik und Dynamik, PTB-Braunschweig
Gazioch, Susanne; 1.7, Akustik und Dynamik, PTB-Braunschweig
Quelle/Jahr Metrologia: 53 (2016), 3, 986 - 990
ISSN 0026-1394 (PRINT) ; 1681-7575 (ONLINE)
DOI
URL
Verlag Bristol: IOP
Zusammenfassung This paper describes the influence of a constant curvature of the shaker's air bearing guidance on a low frequency primary accelerometer calibration. Based on the mathematical model, three different methods are developed that allow a quantitative evaluation of this disturbing effect and thus the correction of the resulting systematic deviation. All three methods are applied to the example of a primary low frequency accelerometer calibration performed at PTB, and the results are given in comparison to the original uncorrected magnitude of sensitivity results.

Zitierung

Bruns, T. & Gazioch, S. (2016). Correction of shaker flatness deviations in very low frequency primary accelerometer calibration. Metrologia, 53(3), 986–990. https://doi.org/10.1088/0026-1394/53/3/986