Zugriffsnummer 36023
Dokumenttyp Zeitschriftenartikel
Peer Review unbekannt
Sprache Englisch
Titel Bilateral comparison in primary angular vibration calibration CCAUV.V-S1
Autor(in); Institution
Täubner, Angelika; 1.7, Akustik und Dynamik, PTB-Braunschweig
Bruns, Thomas; 1.7, Akustik und Dynamik, PTB-Braunschweig
Quelle/Jahr Metrologia: 51 (2014), 1A, Tech. Suppl., 1 - 53
Artikelnummer 09006
Availability [online only]
ISSN 0026-1394 (PRINT) ; 1681-7575 (ONLINE)
DOI
Verlag Bristol: IOP Publishing
Zusammenfassung A comparison CCAUV.V-S1 was organized to compare measurements of sinusoidal angular accelerations in the frequency range from 0.4 Hz to 1 kHz. This was a bilateral comparison between the Korea Research Institute of Standards and Science (KRISS), Republic of Korea, and the Physikalisch-Technische Bundesanstalt (PTB), Germany. Both NMIs applied laser interferometry in compliance with ISO 16063-15:2006 “Methods for the calibration of vibration and shock transducers - Part 15: Primary angular vibration calibration by laser interferometry”. The complex voltage sensitivity (magnitude and phase) of one angular transfer standard in the frequency range of 0.4 Hz to 100 Hz and the complex charge sensitivity (magnitude and phase) of one angular reference standard in the frequency range of 1 Hz to 1000 Hz were measured

Zitierung

Täubner, A. & Bruns, T. (2014). Bilateral comparison in primary angular vibration calibration CCAUV.V-S1. Metrologia, 51(1A, Tech. Suppl.), 1–53. https://doi.org/10.1088/0026-1394/51/1a/09006

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag