Titel Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities EURAMET.RI(I)-S4.1)
Autor(in); Institution
Csete, I.; International Atomic Energy Agency (IAEA), Vienna, AUSTRIA
Büermann, L.; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig
Gomola, I.; International Atomic Energy Agency (IAEA), Vienna, AUSTRIA
Quelle/Jahr Metrologia: 53 (2016), Tech. Suppl., 1 - 14
Artikelnummer 06005
Availability [online only]
ISSN 0026-1394 (PRINT) ; 1681-7575 (ONLINE)
Verlag Bristol: IOP
Freie Schlagworte air kerma ; mammographie x-ray qualities ; comparison
Zusammenfassung A comparison of the air kerma standards for x-radiation qualities used in mammography was performed  between  the  PTB and  the  IAEA.  Two  reference-class  ionization  chamber  types Radcal RC6M and Magna A650 of the IAEA and tungsten anode based beam qualities with Mo and Al external filtrations (W+Mo, W+Al) established at both laboratories were selected for  the  comparison.  The  calibration  coefficients,  NK_air ,  were  determined  for  the  transfer chambers  at  the  PTB  in  May  2015  and  before  and  after  this  at  the  IAEA  Dosimetry Laboratory.  The  results  show  good  agreement,  to  be  well  within  the  0.55 %  standard uncertainty of the comparison. Correction factors to determine N K_air  for these beam qualities based on calibration in RQR-M mammography beam qualities, established according to the IEC 61267 standard, were also calculated for the Radcal RC6M, 10X5-6M, and Magna A650 types of chambers.

Zitierung

Csete, I., Büermann, L., & Gomola, I. (2016). Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities EURAMET.RI(I)-S4.1). Metrologia, 53(Tech. Suppl.), 1–14.

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag