| Titel | Portable E-field strength meter and its traceable calibration up to 1 GHz using a “μTEM” cell |
| Autor(in); Institution |
Münter, Klaus; 2.21, Elektromagnetische Felder, PTB-Braunschweig
Pape, Reiner; 2.21, Elektromagnetische Felder, PTB-Braunschweig
Glimm, Jochen; 2.21, Elektromagnetische Felder, PTB-Braunschweig
|
| Quelle/Jahr | IEEE Transactions on Instrumentation and Measurement: 46 (1997), 2, 549 - 550 |
| ISSN | 0018-9456 |
| DOI | |
| Verlag | New York, NY: IEEE |
| Freie Schlagworte | calibration ; electromagnetic compatibility ; field strength measurement ; frequency response ; measurement standards ; test facilities |
| Zusammenfassung | The newly developed E-field meter is intended as a transfer instrument between a calibration laboratory and an EMC test facility to measure rf fields that are traceable to national standards. The system is based on a flat miniature E-field sensor and uses software corrections for frequency response, linearity and temperature effects. The calibration procedure with a special “μ-TEM” cell is described and an uncertainty budget is given |