Titel Final report on key comparison EURAMET.M.P-K13 in the range 50 MPa to 500 MPa of hydraulic gauge pressure
Autor(in); Institution
Kocas, I.; National Metrology Institute of Turkey, Ankara, TURKEY
Sabuga, Wladimir; 3.3, Thermophysikalische Größen, PTB-Braunschweig
Bergoglio, M.; Istituto Nazionale di Ricerca Metrologica (INRIM), Torino, ITALY
Eltaweel, A.; National Institute for Standards (NIS), Giza, EGYPT
Korasie, C.; National Metrology Institute of South Africa (NMISA), Pretoria, SOUTH AFRICA
Farar, P.; Slovak Institute of Metrology (SMU), Bratislava, SLOVAKIA
Setina, J.; Institute of Metals and Technology (IMT), Kolkata, SLOVENIA
Waller, B.; National Physical Laboratory (NPL), Teddington, UK
Durgut, Y.; National Metrology Institute of Turkey, Ankara, TURKEY
Quelle/Jahr Metrologia: 52 (2015), Tech. Suppl., 43 S.
Artikelnummer 07008
Availability [online only]
ISSN 0026-1394 (PRINT) ; 1681-7575 (ONLINE)
Verlag Bristol [u.a.]: IOP [u.a.]
Zusammenfassung The regional key comparison EURAMET.M.P-K13 for pressure measurements in liquid media from 50 MPa to 500 MPa was piloted by the TÜBITAK UME Pressure Group Laboratories, Turkey. The transfer standard was a DH-Budenberg pressure balance with a free deformation piston-cylinder unit of 2 mm² nominal effective area. Six laboratories from the EURAMET region, namely PTB, INRIM, SMU, IMT, NPL and UME, and two laboratories from the AFRIMETS region, NIS and NMISA, participated in the comparison. Participant laboratories and countries are given in the bottom of the page. PTB participated in this comparison to provide a link to corresponding 500 MPa CCM key comparison CCM.P-K13. The results of all participants excepting NMISA and NPL were found to be consistent with the reference value of the actual comparison and of CCM.P-K13 within their claimed uncertainties (k = 2), at all pressures. Compared in pairs all laboratories with exception of NPL and NMISA demonstrate their agreement with each other within the expanded uncertainties (k = 2) at all pressures. The results are therefore considered to be satisfactory.