Zusammenfassung The emission of a full body scanner Rohde & Schwarz QPS 100TM has been measured traceable to the SI units in the operation frequency range of the device between 70 GHz and 80 GHz. The measurements which have been performed with a calibrated spectrum analyzer and calibrated horn antennas are discussed with regard to the safety limits for non-ionizing electromagnetic radiation.