| Zugriffsnummer | 34809 |
| Dokumenttyp | Konferenzartikel |
| Peer Review | unbekannt |
| Sprache | Englisch |
| Titel | Emission measurement of a full body mm wave scanner |
| Autor(in); Institution |
Kleine-Ostmann, Thomas; 2.2, Hochfrequenz und Felder, PTB-Braunschweig
Schrader, Thorsten; 2.2, Hochfrequenz und Felder, PTB-Braunschweig
|
| Quelle/Jahr | Proceedings of the 40th International Conference on Infrared, Millimeter, and Terahertz Waves:(2015), 2 S. |
| Availability | [USB-Stick] |
| DOI | |
| Konferenzangaben | 40th International Conference on Infrared, Millimeter, and Terahertz Waves, Hong Kong, 23-28, August, 2015, China |
| Freie Schlagworte | millimeter waves ; emission measurements ; full body scanner |
| Zusammenfassung | The emission of a full body scanner Rohde & Schwarz QPS 100TM has been measured traceable to the SI units in the operation frequency range of the device between 70 GHz and 80 GHz. The measurements which have been performed with a calibrated spectrum analyzer and calibrated horn antennas are discussed with regard to the safety limits for non-ionizing electromagnetic radiation. |