Zugriffsnummer 34804
Dokumenttyp Zeitschriftenartikel
Peer Review unbekannt
Sprache Englisch
Titel Validation of a quantized-current source with 0.2 ppm uncertainty
Autor(in); Institution
Stein, Friederike; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Drung, Dietmar; 7.2, Kryophysik und Spektrometrie, PTB-Berlin
Fricke, Lukas; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Scherer, Hansjörg; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Hohls, Frank; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Leicht, Christoph; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Götz, Martin; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Krause, Christian; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Behr, Ralf; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Pesel, Eckart; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Pierz, Klaus; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Siegner, Uwe; 2, Elektrizität, PTB-Braunschweig
Ahlers, Franz Josef; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Schumacher, Hans Werner; 2.5, Halbleiterphysik und Magnetismus, PTB-Braunschweig
Quelle/Jahr Applied Physics Letters: 107 (2015), 10, 103501-1 - 103501-5
ISSN 1077-3118 (PRINT) ; 0003-6951 (ONLINE)
DOI
URL
Verlag College Park, Md.: American Institute of Physics
Freie Schlagworte Electron Transfer ; Error analysis ; Quantum dots ; Quantum measurement theory ; Time measurement
Zusammenfassung We report on high-accuracy measurements of quantized current, sourced by a tunable-barrier single-electron pump at frequencies f up to 1 GHz. The measurements were performed with an ultrastable picoammeter instrument, traceable to the Josephson and quantum Hall effects. Current quantization according to I = ef with e being the elementary charge was confirmed at f = 545 MHz with a total relative uncertainty of 0.2 ppm, improving the state of the art by about a factor of 5. The accuracy of a possible future quantum current standard based on single-electron transport was experimentally validated to be better than the best (indirect) realization of the ampere within the present SI.

Zitierung

Stein, F., Drung, D., Fricke, L., Scherer, H., Hohls, F., Leicht, C., Götz, M., Krause, C., Behr, R., Pesel, E., Pierz, K., Siegner, U., Ahlers, F. J., & Schumacher, H. W. (2015). Validation of a quantized-current source with 0.2 ppm uncertainty. Applied Physics Letters, 107(10), 103501-1–103501-5. https://doi.org/10.1063/1.4930142