Zugriffsnummer 34390
Dokumenttyp Konferenzartikel
Peer Review unbekannt
Sprache Englisch
Titel A system for in situ S-parameter measurements of MR transmit arrays
Autor(in); Institution
Weidemann, Gerd; 8.1, Medizinische Messtechnik, PTB-Berlin
Seifert, Frank; 8.1, Medizinische Messtechnik, PTB-Berlin
Hoffmann, Werner; 8.1, Medizinische Messtechnik, PTB-Berlin
Seemann, Reiner; 8.1, Medizinische Messtechnik, PTB-Berlin
Waxmann, Patrick; 8.1, Medizinische Messtechnik, PTB-Berlin
Ittermann, Bernd; 8.1, Medizinische Messtechnik, PTB-Berlin
Quelle/Jahr ISMRM 23th Annual Meeting & Exhibition: 30 May - 5 June, 2015; Toronto, Canada:(2015), 1 S.
Availability [CD-ROM] ; file name: 1775.pdf
ISSN 1545-4428
Verlag Berkeley, Calif.: ISMRM
Konferenzangaben 23th ISMRM Annual Meeting & Exhibition, Toronto, 30 May – 5 June, 2015, Canada
Zusammenfassung The  S-parameter  matrix  of  the  3T  pTx  coil    shows  low  direct reflection but strong coupling to the next nearest neighbor (nnn) elements. Since the coil is not shielded it is quite sensitive to its surrounding. Placed inside a 38-cm shim insert and a few centimeters off axis, the direct reflection is increased and the coupling to the nnn elements is reduced. The elements closer to the shim insert become more detuned and a reflected amplitude of up to 0.59 (35% reflected power) is measured directly in the transmit channel. The power accepted by the coil is even lower than 65% because of additional losses in cables and Tx/Rx switch. The S-parameter matrix of the shielded 7T array shows a different characteristic. The direct reflection in the transmit channel is higher but the coupling decreases continuously with the distance of the element.

Zitierung

Weidemann, G., Seifert, F., Hoffmann, W., Seemann, R., Waxmann, P., & Ittermann, B. (2015). A system for in situ S-parameter measurements of MR transmit arrays. 23th ISMRM Annual Meeting & Exhibition, Toronto, 30 May – 5 June, 2015, Canada.