| Titel | The practical peak voltage of diagnostic X-ray generators |
| Autor(in); Institution |
Kramer, Hans-Michael; 6.21, Dosimetrie der Röntgenstrahlung, PTB-Braunschweig
Selbach, Hans-Joachim; 6.21, Dosimetrie der Röntgenstrahlung, PTB-Braunschweig
Iles, W. J.; National Radiological Protection Board, Didcot, Oxon, UK
|
| Quelle/Jahr | British Journal of Radiology: 71 (1998), 842, 200 - 209 |
| ISSN | 0007-1285 |
| DOI | |
| Verlag | London: British Institute of Radiology |
| Zusammenfassung | A new quantity termed the "practical peak voltage" is proposed. This quantity is derived by equating the low level contrast in an exposure made with an X-ray tube connected to a generator delivering any arbitrary wave form, to the contrast produced by the same X-ray tube connected to a constant potential generator. Out of the great number of possible contrast configurations one is selected as being suitable for diagnostic radiology. By means of an eigenvalue problem a direct link is established between the electrical quantity X-ray tube voltage and the practical peak voltage which was initially defined through the properties of the X-ray field. It is shown that the spread in total X-ray tube filtration as encountered in medical diagnostic radiology can influence the result of a measurement of the practical peak voltage only marginally. |