| Zugriffsnummer | 33723 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | H*(10) Response of the REM-2 Recombination Chamber in Monoenergetic Neutron Fields |
| Autor(in); Institution |
Golnik, N.; Institute of Atomic Energy (IAE), Otwock-Swierk, POLAND
Brede, Hein Jürgen; 6.31, Ionendosimetrie und Beschleunigerbetrieb, PTB-Braunschweig
|
| Quelle/Jahr | Radiation Protection Dosimetry: 74 (1997), 3, 139 - 144 |
| ISSN | 0144-8420 (PRINT) ; 1742-3406 (ONLINE) |
| DOI | |
| Verlag | Oxford: Oxford Univ. Press |
| Zusammenfassung | Measurements with the REM-2 recombination chamber were performed in the monoenergetic reference neutron fields at the PTB. The energy dependence of the H*(10) ambient dose equivalent response was investigated for 12 different neutron energies ranging from 75 keV to 19 MeV. A new simplified data analysis for the determination of H*(10) with the ICRP 60 quality factor was tested. The results indicate that the REM-2 recombination chamber is well suited for the direct determination of the ambient dose equivalent in mixed neutron/photon radiation fields. |