Zugriffsnummer 33723
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel H*(10) Response of the REM-2 Recombination Chamber in Monoenergetic Neutron Fields
Autor(in); Institution
Golnik, N.; Institute of Atomic Energy (IAE), Otwock-Swierk, POLAND
Brede, Hein Jürgen; 6.31, Ionendosimetrie und Beschleunigerbetrieb, PTB-Braunschweig
Quelle/Jahr Radiation Protection Dosimetry: 74 (1997), 3, 139 - 144
ISSN 0144-8420 (PRINT) ; 1742-3406 (ONLINE)
DOI
Verlag Oxford: Oxford Univ. Press
Zusammenfassung Measurements with the REM-2 recombination chamber were performed in the monoenergetic reference neutron fields at the PTB. The energy dependence of the H*(10) ambient dose equivalent response was investigated for 12 different neutron energies ranging from 75 keV to 19 MeV. A new simplified data analysis for the determination of H*(10) with the ICRP 60 quality factor was tested. The results indicate that the REM-2 recombination chamber is well suited for the direct determination of the ambient dose equivalent in mixed neutron/photon radiation fields.

Zitierung

Golnik, N. & Brede, H. J. (1997). H*(10) Response of the REM-2 Recombination Chamber in Monoenergetic Neutron Fields. Radiation Protection Dosimetry, 74(3), 139–144. https://doi.org/10.1093/oxfordjournals.rpd.a032189

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag