| Zugriffsnummer | 33546 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | Wavelength measurement of a 544-nm FM-I2-stabilised He-Ne-Laser |
| Autor(in); Institution |
Bönsch, Gerhard; 5.13, Interferentielle Längenmessung, PTB-Braunschweig
Nicolaus, Arnold; 5.13, Interferentielle Längenmessung, PTB-Braunschweig
Brand, Uwe; 5.11, Maskenmesstechnik, PTB-Braunschweig
|
| Quelle/Jahr | Optik: 107 (1998), 3, 127 - 131 |
| ISSN | 0030-4026 |
| Verlag | Jena: Fischer |
| Zusammenfassung | The vacuum wavelength of a 544 nm iodine-stabilised He-Ne laser was measured by means of the PTB's Michelson interferometer, which had been designed for highly accurate wavelength comparisons. One of the two longitudinal models of the internal mirror He-Ne laser was selected and frequency stabilised using FM saturation spectroscopy at iodine by means of an external cell. Reference wavelengths were obtained from an I2-stabilised He-Ne laser at 633 nm and an 12-stabilised Ar+-laser at 515 nm. The wavelength of the 544nm 12-stabilised He-Ne laser was found to be λ = 543,51633304 nm with a relative standard uncertainty of 8,4. 10-11 |