Titel A new FEL-type quartz-halogen lamp as an improved standard of spectral irradiance
Autor(in); Institution
Metzdorf, Jürgen; 4.1, Licht und Strahlung, PTB-Braunschweig
Sperling, A.; PTB-Braunschweig
Winter, Stefan; 4.13, Angewandte Radiometrie, PTB-Braunschweig
Raatz, Karl-Heinz; 4.11, Radiometrische Einheiten, PTB-Braunschweig
Möller, Werner; 4.13, Angewandte Radiometrie, PTB-Braunschweig
Quelle/Jahr Metrologia: 35 (1998), 4, 423 - 426
ISSN 0026-1394 (PRINT) ; 1681-7575 (ONLINE)
DOI
Verlag Bristol [u.a.]: Institute of Physics Publishing (IOP) ; Bureau International des Poids et Mesures
Konferenzangaben 6th International Conference on Radiometry, NEWRAD '97, Tucson, Ariz., 27-29, October, 1997, USA
Klassifikationscode PACS: 42.72.-g ; PACS: 06.20.F- ; PACS: 07.60.Dq
Freie Schlagworte Instrumentation and measurement ; Optics, quantum optics and lasers
Zusammenfassung Based on extensive investigations of almost 100 high-power quartz-halogen lamps covering thirteen different types, seven different manufacturers and different batches with a total operating time of about 27000 h, a new type of lamp (1000 W T6 modified FEL-type lamp, Osram Sylvania Inc.) has been selected and its operating conditions have been optimized. This lamp, which after seasoning is about ten times more stable than the best lamps used before, is recommended as a reference and transfer standard of spectral irradiance in the whole range between 250 nm and 3 µm. While excellent short-term and long-term stability over operating times of several hundreds of hours can be achieved, the disadvantages of the new standard lamp lie in a long seasoning time (80 h to 100 h) and a noticeable sensitivity of the seasoned lamps to shock and vibrations. Typical results of the characterization, calibration and reproducibility of the new standard lamp are presented and discussed in order to illustrate its performance and to demonstrate how to use it under optimal operation conditions.

Zitierung

Metzdorf, J., Sperling, A., Winter, S., Raatz, K.-H., & Möller, W. (1998). A new FEL-type quartz-halogen lamp as an improved standard of spectral irradiance. Metrologia, 35(4), 423–426. https://doi.org/10.1088/0026-1394/35/4/35