Zugriffsnummer 33467
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Influence of eccentric mounting on the calibration of torque measuring devices
Autor(in); Institution
Brüge, Andreas; 1.22, Drehmoment, PTB-Braunschweig
Quelle/Jahr Accuracy, assurance in force, torque and mass measurements : basic metrology to industrial applications ; proceedings of the 15th international conference:(1996), 255 - 259
Herausgeber(in)
Robles Carbonell, José A.
Verlag Madrid: Centro Espanol de Metrologia
Konferenzangaben International Conference Accuracy, Assurance in Force, Torque and Mass Measurements, Madrid, 07-11, November, 1996, Spain
Zusammenfassung In the calibration of force transducers high tolerances for eccentric mounting of the transducer are possible because flexible couplings are used. In contrast to this, exact mounting is required for the calibration of torque wrenches, since a deviation of the wrench from its centric position alters the effective torque applied to it.

Zitierung

Brüge, A. (1996). Influence of eccentric mounting on the calibration of torque measuring devices. International Conference Accuracy, Assurance in Force, Torque and Mass Measurements, Madrid, 07-11, November, 1996, Spain.