| Zugriffsnummer | 33467 |
| Dokumenttyp | Konferenzartikel |
| Sprache | Englisch |
| Titel | Influence of eccentric mounting on the calibration of torque measuring devices |
| Autor(in); Institution |
Brüge, Andreas; 1.22, Drehmoment, PTB-Braunschweig
|
| Quelle/Jahr | Accuracy, assurance in force, torque and mass measurements : basic metrology to industrial applications ; proceedings of the 15th international conference:(1996), 255 - 259 |
| Herausgeber(in) |
Robles Carbonell, José A.
|
| Verlag | Madrid: Centro Espanol de Metrologia |
| Konferenzangaben | International Conference Accuracy, Assurance in Force, Torque and Mass Measurements, Madrid, 07-11, November, 1996, Spain |
| Zusammenfassung | In the calibration of force transducers high tolerances for eccentric mounting of the transducer are possible because flexible couplings are used. In contrast to this, exact mounting is required for the calibration of torque wrenches, since a deviation of the wrench from its centric position alters the effective torque applied to it. |