Titel Response of the alanine/ESR dosimeter to radiation from an Ir-192 HDR brachytherapy source
Autor(in); Institution
Anton, Mathias; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig
Hackel, Thomas; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig
Zink, K.; IMPS, University of Applied Sciences Gießen-Friedberg, Institut für Medizinische Physik und Strahlenschutz, Gießen, GERMANY
Voigts-Rhetz, P. von; IMPS, University of Applied Sciences Gießen-Friedberg, Institut für Medizinische Physik und Strahlenschutz, Gießen, GERMANY
Selbach, Hans-Joachim; PTB-Braunschweig, retired staff
Quelle/Jahr Physics in Medicine and Biology: 60 (2014), 1, 175 - 193
ISSN 0031-9155
DOI
Verlag Bristol: IOP
Klassifikationscode PACS: 87.53.Jw ; PACS: 87.55.K- ; PACS: 87.80.-y ; PACS: 87.55.Gh
Freie Schlagworte Wasser-Energiedosis ; Alanin ; ESR ; EPR ; Sekundärnormal-Messsystem ; Ansprechvermögen ; Iridium ; Ir-192 ; Brachytherapie ; HDR
Zusammenfassung The response of the alanine dosimeter to radiation from an Ir-192 source with respect to the absorbed dose to water, relative to Co-60 radiation, was determined experimentally as well as by Monte Carlo simulations. The experimental and Monte Carlo results for the response agree well within the limits of uncertainty. The relative response decreases with an increasing distance between the measurement volume and the source from approximately 98 % at a 1 cm distance to 96 % at 5 cm. The present data are more accurate, but agree well with data published by Schaeken et al. (2011). The decrease of the relative response with an increasing distance that had already been observed by these authors is confirmed. In an appendix, the properties of the alanine dosimeter with respect to volume and sensitivity corrections are investigated.The inhomogeneous distribution of the detection probability that was taken into account for the analysis was determined experimentally.

Zitierung

Anton, M., Hackel, T., Zink, K., Voigts-Rhetz, P. V., & Selbach, H.-J. (2014). Response of the alanine/ESR dosimeter to radiation from an Ir-192 HDR brachytherapy source. Physics in Medicine and Biology, 60(1), 175–193. https://doi.org/10.1088/0031-9155/60/1/175