Titel Influence of the phantom shape (slab, cylinder or Alderson) on the performance of an Hp(3) eye dosemeter
Autor(in); Institution
Behrens, Rolf; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Hupe, Oliver; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Quelle/Jahr Radiation Protection Dosimetry: 168 (2016), 4, 441 - 449
ISSN 0144-8420 (PRINT) ; 1742-3406 (ONLINE)
DOI
URL
Verlag Oxford: OUP
Zusammenfassung In the past, the operational quantity Hp(3) was defined for calibration purposes in a slab phantom. Recently, an additional phantom in the form of a cylinder has been suggested for eye lens dosimetry, as a cylinder much better approximates the shape of a human head. Therefore, this work investigates which of the two phantoms, slab or cylinder, is more suitable for calibrations and type tests of eye dosemeters. For that purpose, a typical Hp(3) eye dosemeter was irradiated on a slab, a cylinder and on a human-like Alderson phantom. It turned out that the response on the three phantoms is nearly equal for angles of radiation incidence up to 458 and deviates only at larger angles of incidence. Thus, calibrations (usually performed at 08 radiation incidence) are practically equivalent on both the slab and the cylinder phantoms. However, type tests (up to 758 or even 908 radiation incidence) should be carried out on a cylinder phantom, as also for large angles of incidence the response on the cylinder and the Alderson phantoms is rather similar, whereas the response on the slab significantly deviates from the one on the Alderson phantom.

Zitierung

Behrens, R. & Hupe, O. (2016). Influence of the phantom shape (slab, cylinder or Alderson) on the performance of an Hp(3) eye dosemeter. Radiation Protection Dosimetry, 168(4), 441–449. https://doi.org/10.1093/rpd/ncv366

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag