Zugriffsnummer 33010
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel Calibration of a fiber tip ultrasonic sensor up to 50 MHz and the application to shock wave measurement
Autor(in); Institution
Koch, Christian; 1.43, Ultraschall, PTB-Braunschweig
Ludwig, Gerhard; 1.43, Ultraschall, PTB-Braunschweig
Molkenstruck, Walter; 1.43, Ultraschall, PTB-Braunschweig
Quelle/Jahr Ultrasonics: 36 (1998), 1-5, 721 - 725
ISSN 0041-624X
DOI
Verlag Amsterdam [u.a.]: Elsevier
Freie Schlagworte Fiber-optic sensors ; Hydrophone calibration ; Interferometry ; Ultrasound measurement
Zusammenfassung An optical calibration technique is described, which allows the investigation of the frequency response of ultrasonic detectors up to a frequency of 50 MHz. This technique is applied to determine the transfer function of a metal-coated ultrasonic fiber tip sensor which is based on an interferometric measurement of the displacement using a Michelson interferometer. The transfer function is used to correct shock wave data which are, by this means, measured with a calibrated system.

Zitierung

Koch, C., Ludwig, G., & Molkenstruck, W. (1998). Calibration of a fiber tip ultrasonic sensor up to 50 MHz and the application to shock wave measurement. Ultrasonics, 36(1-5), 721–725. https://doi.org/10.1016/s0041-624x(97)00121-2