| Zugriffsnummer | 33010 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | Calibration of a fiber tip ultrasonic sensor up to 50 MHz and the application to shock wave measurement |
| Autor(in); Institution |
Koch, Christian; 1.43, Ultraschall, PTB-Braunschweig
Ludwig, Gerhard; 1.43, Ultraschall, PTB-Braunschweig
Molkenstruck, Walter; 1.43, Ultraschall, PTB-Braunschweig
|
| Quelle/Jahr | Ultrasonics: 36 (1998), 1-5, 721 - 725 |
| ISSN | 0041-624X |
| DOI | |
| Verlag | Amsterdam [u.a.]: Elsevier |
| Freie Schlagworte | Fiber-optic sensors ; Hydrophone calibration ; Interferometry ; Ultrasound measurement |
| Zusammenfassung | An optical calibration technique is described, which allows the investigation of the frequency response of ultrasonic detectors up to a frequency of 50 MHz. This technique is applied to determine the transfer function of a metal-coated ultrasonic fiber tip sensor which is based on an interferometric measurement of the displacement using a Michelson interferometer. The transfer function is used to correct shock wave data which are, by this means, measured with a calibrated system. |