Zugriffsnummer 31752
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Suitability of the echo-time-shift method as laboratory standard for thermal ultrasound dosimetry
Autor(in); Institution
Fuhrmann, Tina; University of Applied Sciences, Merseburg, GERMANY
Georg, Olga; 1.6, Schall, PTB-Braunschweig
Haller, Julian; 1.6, Schall, PTB-Braunschweig
Jenderka, Klaus-Vitold; University of Applied Sciences, Merseburg, GERMANY
Quelle/Jahr 14th International Symposium on Therapeutic Ultrasound 2014:(2017), 150005-1 - 150005-5
Schriftenreihe AIP conference proceedings: 1821
ISBN 978-0-7354-1489-1
DOI
Verlag Melville, NY: AIP
Konferenzangaben 14th International Symposium on Therapeutic Ultrasound (ISTU14), Las Vegas, Nev., 02-05, April, 2014, USA
Klassifikationscode PACS: 87.50.Yk ; PACS: 42.80.Ev ; PACS: 43.80.V
Freie Schlagworte Therapeutischer Ultraschall ; HITU ; diagnostischer Ultraschall ; Temperaturmessung ; ultrasound therapy ; non-invasive thermometry ; HIFU ; dosimetry ; metrology
Zusammenfassung Ultrasound therapy is a promising, non-invasive application with potential to significantly improve cancer therapies like surgery, viro- or immunotherapy. This therapy needs faster, cheaper and more easy-to-handle quality assurance tools for therapy devices as well as possibilities to verify treatment plans and for dosimetry. This limits comparability and safety of treatments. Accurate spatial and temporal temperature maps could be used to overcome these shortcomings. In this contribution first results of suitability and accuracy investigations of the echo-time-shift method for two-dimensional temperature mapping during and after sonication are presented. The analysis methods used to calculate time-shifts were a discrete frame-to-frame and a discrete frame-to-base-frame algorithm as well as a sigmoid fit for temperature calculation. In the future accuracy could be significantly enhanced by using continuous methods for time-shift calculation. Further improvements can be achieved by improving filtering algorithms and interpolation of sampled diagnostic ultrasound data. It might be a comparatively accurate, fast and affordable method for laboratory and clinical quality control.

Zitierung

Fuhrmann, T., Georg, O., Haller, J., & Jenderka, K.-V. (2017). Suitability of the echo-time-shift method as laboratory standard for thermal ultrasound dosimetry. 14th International Symposium on Therapeutic Ultrasound (ISTU14), Las Vegas, Nev., 02-05, April, 2014, USA. https://doi.org/10.1063/1.4977649

Exportieren

PTB-Publica MenĂ¼

Sprache wechseln: uk flag