Zugriffsnummer 31646
Dokumenttyp Zeitschriftenartikel
Peer Review unbekannt
Sprache Englisch
Titel COOMET regional comparison of national measurement standards of air kerma for Cs-137- gamma-radiation at protection level
Autor(in); Institution
Büermann, Ludwig; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig
Oborin, A. V.; D.I. Mendeleyev Institute for Metrology (VNIIM), St. Petersburg, RUSSIA
Milevsky, V. S.; Bellorussian State Institute for Metrology (BelGIM), Minsk, REPUBLIC OF BELARUS
Walwyn Salas, Gonzalo; Centro de Protección e Higiene de las Radiaciones (CPHR), Habana, CUBA
Sukhishvili, S.; Georgian National Agency for Standards, Technical Regulations and Metrology, Institute of Metrology (GEOSTM), Tbilisi, GEORGIA
Ginga, Ion; National Institute of Standardization and Metrology (INSM), Chisinau, REPUBLIC OF MOLDOVA
Ivanov, R.; Bulgarian Institute of Metrology (BIM), Lovech, BULGARIA
Gudelis, A.; Metrology Department Center for Physical Sciences and Technology (VMT/FTMC), Vilnius, LITHUANIA
Gomola, Igor; International Atomic Energy Agency (IAEA), Vienna, AUSTRIA
Quelle/Jahr Metrologia: 51 (2014), Tech. Suppl., 1 - 12
Artikelnummer 06005
Availability [online only]
ISSN 0026-1394
DOI
Verlag Bristol: IOP Publishing
Freie Schlagworte Air kerma standards ; Air Kerma standard comparison ; Cs-137 gamma radiation ; COOMET supplementary comparison
Zusammenfassung Results are presented of the COOMET supplementary comparison of the national measurement standards for air kerma in Cs-137 gamma-radiation at protection level (~10 mGy/h). Ten National Metrology Institutes from the COOMET organization and the International Atomic Energy Agency participated in this COOMET project no. 445. The PTB acted as pilot laboratory. Two of the participants, the SMU (Slovakia) and the NSC-"IM" (Ukraine) participated in the measurements but did not submit a valid report of results. The comparison reference value (CRV) was obtained as the mean result of the PTB and the VNIIM, both of which had previously taken part in the key comparison BIPM-RI(I)-K5. The degree of equivalence with the CRV was evaluated. The results were consistent within the relative standard uncertainties of the comparison ranging from 0.28% – 1.3% and deviated from the CRV by less than 1%.

Zitierung

Büermann, L., Oborin, A. V., Milevsky, V. S., Walwyn Salas, G., Sukhishvili, S., Ginga, I., Ivanov, R., Gudelis, A., & Gomola, I. (2014). COOMET regional comparison of national measurement standards of air kerma for Cs-137- gamma-radiation at protection level. Metrologia, 51(Tech. Suppl.), 1–12. https://doi.org/10.1088/0026-1394/51/1a/06005

Exportieren