| Zugriffsnummer | 30656 |
| Dokumenttyp | Zeitschriftenartikel |
| Peer Review | unbekannt |
| Sprache | Englisch |
| Titel | Final report of key comparison CCM.P-K12 for very low helium flow rates (leak rates) |
| Autor(in); Institution |
Jousten, Karl; 7.5, Wärme und Vakuum, PTB-Berlin
Aria, K.; National Metrology Institute of Japan, Tsukuba, Ibaraki, JAPAN
Becker, Ute; 7.5, Wärme und Vakuum, PTB-Berlin
Bodnar, Olha; 8.4, Mathematische Modellierung und Datenanalyse, PTB-Berlin
Boineau, F.; Laboratoire national de métrologie et d'eessais, Paris, FRANCE
Fedchak, J.A.; National Institute of Standards and Technology, Gaithersburg, USA
Gorobey, V.; D.I. Mendeleyev Institute VNIIM, St. Petersburg, RUSSIA
Jian, Wu; National Metrology Centre, SINGAPORE
Mari, D.; Istituto Nazionale di Ricerca Metrologica, Torino, ITALY
Mohan, P.; CSIR-National PHysical Maboratory India, New Delhi, INDIA
Setina, J.; Institute of Metals and Technology, Ljubljana, SLOVENIA
Toman, B.; National Institute of Standards and Technology, Gaithersburg, USA
Vicar, M.; Czech Metrological Institute, Brno, CZECH REPUBLIC
Yan, Yu Hong; National Institute of Metrology, San Han Dong Lu Beijing, CHINA
|
| Quelle/Jahr | Metrologia: 50 (2013), 1A, Tech. Suppl., 50 S. |
| Artikelnummer | 07001 |
| Availability | [online only] |
| ISSN | 0026-1394 (PRINT) ; 1681-7575 (ONLINE) |
| DOI | |
| Verlag | Bristol [u.a.]: IOP [u.a.] |
| Freie Schlagworte | vacuum technology ; helium ; leak rate |
| Zusammenfassung | Quantitative leak tests with vacuum technology have become an important tool in industry for safety and operational reasons and to meet environmental regulations. In the absence of a relevant key comparison, so far, there are no calibration measurement capabilities published in the BIPM data base. To enable national metrology institutes providing service for leak rate calibrations to apply for these entries in the data base and to ensure international equivalence in this field, key comparison CCM.P-K12 was organized. The goal of this comparison was to compare the national calibration standards and procedures for helium leak rates. Two helium permeation leak elements of 4×10−11 mol/s (L1) and 8×10−14 mol/s (L2) served as transfer standards and were measured by 11 national metrology institutes for L1 and 6 national metrology institutes for L2. Equivalence could be shown for 8 laboratories in the case of L1 and for all 6 in the case of L2. Three different evaluation methods were applied and are presented in this report, but the random effects model was accepted as most suitable in our case. |
Zitierung
Jousten, K., Aria, K., Becker, U., Bodnar, O., Boineau, F., Fedchak, J., Gorobey, V., Jian, W., Mari, D., Mohan, P., Setina, J., Toman, B., Vicar, M., & Yan, Y. H. (2013). Final report of key comparison CCM.P-K12 for very low helium flow rates (leak rates). Metrologia, 50(1A, Tech. Suppl.), 50 S. https://doi.org/10.1088/0026-1394/50/1a/07001