Zugriffsnummer 30559
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel High-current hollow-cathode source as a radiant intensity standard in the 40–125-nm wavelength range
Autor(in); Institution
Hollandt, Jörg; 9.23, Vakuum-UV-Strahlungsmeßtechnik, PTB-Berlin
Kühne, Michael; 9.1, Wärmemeßtechnik und Vakuumphysik, PTB-Berlin
Wende, Burkhard; 9.2, Thermometrie und Strahlung, PTB-Berlin
Quelle/Jahr Applied Optics: 33 (1994), 1, 68 - 74
ISSN 0003-6935 (PRINT) ; 1539-4522 (ONLINE)
DOI
Verlag Bristol: IOP-Publ.
Freie Schlagworte sources ; radiometry ; vacuum ultraviolet
Zusammenfassung The radiant intensity of VUV emission lines of a high-current hollow-cathode source has been determined for the 40–125-nm spectral range. The source is operated at a constant current of 1 A with an aluminum cathode. Different rare gases are alternatively used as the buffer gas at pressures of ~ 100 Pa. The radiant intensity has been determined by comparison with the calculable spectral radiant flux of the electron storage ring BESSY. Radiant intensities of the emission lines are in the 7–1400-μW/sr range. The long-term reproducibility of the radiant intensity of the source is within ± 10% (2σ value). The systematic uncertainty of the radiometric calibration is better than 9% (3√ σ value) .

Zitierung

Hollandt, J., Kühne, M., & Wende, B. (1994). High-current hollow-cathode source as a radiant intensity standard in the 40–125-nm wavelength range. Applied Optics, 33(1), 68–74. https://doi.org/10.1364/ao.33.000068

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag