Zugriffsnummer 30508
Dokumenttyp Zeitschriftenartikel
Peer Review unbekannt
Sprache Englisch
Titel A biosensor for L-amino acids using polytyramine for enzyme immobilization
Autor(in); Institution
Cooper, Julia C.; Bayer AG, Leverkusen, GERMANY
Schubert, Florian; 10.12, Messtechnik in der Laboratoriumsmedizin, PTB-Berlin
Quelle/Jahr Electroanalysis: 6 (1994), 11-12, 957 - 961
ISSN 1040-0397 (PRINT) ; 1521-4109 (ONLINE)
DOI
Verlag Weinheim: Wiley-VCH
Freie Schlagworte Biosensor ; Amino acids ; Polytyramine
Zusammenfassung Electrodeposition of polytyramine is demonstrated to be a simple and convenient procedure for electrode modification, generating amine groups to which L-amino acid oxides can be covalently bound. An L-amind acid oxidase (L-AAOD)-polytyramine electrode can be used for detection of L amino acids, via the current due to oxidation of enzymatically produced hydrogen peroxide. The calibration graph of the sensor for phenylalanine is linear up to 1.4 mM with a lower limit of detection of 0.07 mM. The useful measuring range for leucine is between 0.07 and 3 mM. The enzyme-polytyramine electrodes are stable for more than 1 month. The polymer coating affords some protection of the elearode from direct (nonenzymic) oxidation of electroctive amino acids, which may otherwise cause elecatrode fouling, although at present, the polymer selectivity is insufficient to prevent errors in estimation of analyte concentration.

Zitierung

Cooper, J. C. & Schubert, F. (1994). A biosensor for L-amino acids using polytyramine for enzyme immobilization. Electroanalysis, 6(11-12), 957–961. https://doi.org/10.1002/elan.1140061107