| Zugriffsnummer | 30491 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | Improved Cerenkov counting techniques based on a free parameter model |
| Autor(in); Institution |
Kossert, Karsten; 6.1, Radioaktivität, PTB-Braunschweig
Grau Carles, A.; C/Voluntarios Catalanes, Madrid, SPAIN
Nähle, Ole Jens; 6.1, Radioaktivität, PTB-Braunschweig
|
| Quelle/Jahr | Applied Radiation and Isotopes: 86 (2014), 4, 7 - 12 |
| ISSN | 0969-8043 |
| DOI | |
| Verlag | Amsterdam: Elsevier |
| Freie Schlagworte | The TDCR-Cerenkov method was extented and improved ; The model comprises the light dispersion in water ; The wavelength-dependent PMT response is taken into account ; Potential differences of the PMT efficiencies are taken into account |
| Zusammenfassung | In the past few years, two Cerenkov methods were developed to make activity measurements of high-energy beta emitters in liquid scintillation counters with two or three photomultiplier tubes (PMTs) possible. Both methods are based on a free parameter model and make use of the Frank and Tamm theory for the emission of Cerenkov light. In this article, additional effects are discussed and further improvements are presented. Thedependence of the refractive index of water on the wavelength can now be taken into account, which has also an influence on the upper limit of the wavelength region for the production of Cerenkov light. In addtion, the depenceof the PMT response on the wavelengthis taken into account. finally, it is possible to take a potentioal asymmetry of efficiencies in a system with three PMTs in to account. To this end, three free parameters are assigned to each individual PMT and then determined by means of a downhill simplex optimization algorithm. The computedcounting efficiencies for a triple-to-double coincidence ratio (TDCR) system were compared with experimental data for 32P, 89Sr and 90Y. |