Titel Traceable thermoelectric measurements of Seebeck coefficients in the temperature range from 300 K to 900 K
Autor(in); Institution
Lenz, Ernst; 7.4, Temperatur, PTB-Berlin
Edler, Frank; 7.4, Temperatur, PTB-Berlin
Ziolkowski, Pawel; Deutsches Zentrum für Luft- und Raumfahrt e. V., Köln, GERMANY
Quelle/Jahr International Journal of Thermophysics: 34 (2013), 10, 1975 - 1981
ISSN 0195-928X (PRINT) ; 1572-9567 (ONLINE)
DOI
Verlag Dordrecht: Springer
Zusammenfassung This work is focused on the characterization of thermoelectric reference materials with traceable Seebeck coefficients in the temperature range from 300 K to 900 K. The presented measurement system will provide a relative uncertainty of the measurement of the Seebeck coefficient of about 5 %. The Seebeck coefficient represents an important component of the figure of merit ZT and thus the low uncertainty of the Seebeck coefficient will also lower the uncertainty of the ZT value. We also present data which lead to the launch of the certification process of a NiCu-alloy according to the ISO Guide 35.

Zitierung

Lenz, E., Edler, F., & Ziolkowski, P. (2013). Traceable thermoelectric measurements of Seebeck coefficients in the temperature range from 300 K to 900 K. International Journal of Thermophysics, 34(10), 1975–1981. https://doi.org/10.1007/s10765-013-1516-x