Zugriffsnummer 30209
Dokumenttyp Konferenzartikel in Zeitschrift
Sprache Englisch
Titel The double-heatpipe black body: a high-accuracy standard source of spectral irradiance for measurements of T − T90.noitalic
Autor(in); Institution
Stock, Michael; 9.21, Strahlungsthermometrie, PTB-Berlin
Fischer, Joachim; 9.21, Strahlungsthermometrie, PTB-Berlin
Friedrich, Rüdiger; 9.21, Strahlungsthermometrie, PTB-Berlin
Jung, Hans-Joachim; 9.21, Strahlungsthermometrie, PTB-Berlin
Wende, Burkhard; 9.2, Thermometrie und Strahlung, PTB-Berlin
Quelle/Jahr Metrologia: 32 (1995), 6, 441 - 444
ISSN 0026-1394 (PRINT) ; 1681-7575 (ONLINE)
DOI
Verlag Bristol [u.a.]: IOP Publishing
Konferenzangaben 5th International Conference on Radiometry, Berlin, September, 19-21, 1994, Germany
Klassifikationscode PACS: 07.20.Dt ; PACS: 44.40.+a ; PACS: 07.60.Dq ; PACS: 06.20.F- ; PACS: 07.20.Mc
Freie Schlagworte Instrumentation and measurement ; Statistical physics and nonlinear systems
Zusammenfassung The Physikalisch-Technische Bundesanstalt (PTB) has developed a large-area black body (LABB) for accurate realization of the spectral irradiance in the red and near-infrared spectral range. Its temperature is measured by standard Pt-resistance thermometers according to the International Temperature Scale of 1990 (ITS-90). The black body consists of two concentric sodium heatpipes, the inner one with temperature stability of ±2 mK forming the radiating cavity. The relative uncertainty of the spectral irradiance of the LABB in the red spectral range is 6 × 10-4 at the freezing point of silver, the largest contribution resulting from the thermodynamic uncertainty of the temperature scale. This allows the ITS-90 to be checked by a comparison of the LABB with a cryogenic radiometer using calibrated filter radiometers as transfer standards. The thermodynamic temperatures obtained in the range 660 °C to 962 °C agree with T90 within the combined 1σ uncertainties.

Zitierung

Stock, M., Fischer, J., Friedrich, R., Jung, H.-J., & Wende, B. (1995). The double-heatpipe black body: a high-accuracy standard source of spectral irradiance for measurements of T − T₉₀.noitalic. Metrologia, 32(6), 441–444. https://doi.org/10.1088/0026-1394/32/6/7

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag