| Titel | Comparison of the ITS-90 using a transfer standard infrared radiation thermometer between seven EU national metrological institutes |
| Autor(in); Institution |
Machin, G.; National Physical Laboratory, (NPL), Teddington, UK
Ricolfi, T.; Istituto di Metrologia G. Colonnetti, (IMGC), Turin, ITALY
Battuello, M.; Istituto di Metrologia G. Colonnetti, (IMGC), Turin, ITALY
Negro, G.; Institut National de Métrologie, (BNM-INM), Paris, FRANCE
Jung, Hans-Joachim; 9.21, Strahlungsthermometrie, PTB-Berlin
Bloembergen, P.; Nederlands Meetinstituut, Van Swinden Laboratorium (NMi/VSL), Delft, NETHERLANDS
Bosma, R.; Nederlands Meetinstituut, Van Swinden Laboratorium (NMi/VSL), Delft, NETHERLANDS
Ivarsson, J.; Sveriges Provnings och Forskningsinstitut (SP), Borås, SWEDEN
Weckström, T.; Centre for Metrology and Accreditation (CMA), Helsinki, FINLAND
|
| Quelle/Jahr | Metrologia: 33 (1996), 3, 197 - 260 |
| ISSN | 0026-1394 (PRINT) ; 1681-7575 (ONLINE) |
| DOI | |
| Klassifikationscode | PACS: 06.20.F- ; PACS: 07.20.Dt |
| Freie Schlagworte | Instrumentation and measurement |
| Zusammenfassung | The primary purpose of the comparison was to establish the equivalence of the calibration of infrared radiation thermometers within Europe in the temperature range 800 °C to 2000 °C. Seven EU national metrological institutes (NMIs) participated. The equivalence of calibrations was determined by circulating a transfer standard infrared radiation thermometer between the participating laboratories. The calibrations obtained were compared with the reference calibration performed by the co-ordinating laboratory (National Physical Laboratory). It was found that, after correction for the size-of-source effect, the calibrations of the participating laboratories nearly all agree to better than 0,1 % of temperature. |
Zitierung
Machin, G., Ricolfi, T., Battuello, M., Negro, G., Jung, H.-J., Bloembergen, P., Bosma, R., Ivarsson, J., & Weckström, T. (1996). Comparison of the ITS-90 using a transfer standard infrared radiation thermometer between seven EU national metrological institutes. Metrologia, 33(3), 197–260. https://doi.org/10.1088/0026-1394/33/3/2