Autor(in); Institution
Stumper, Ulrich; PTB-Braunschweig, retired staff
Schrader, Thorsten; 2.2, Hochfrequenz und Felder, PTB-Braunschweig
Quelle/Jahr IEEE Transactions on Instrumentation and Measurement: 63 (2014), 6, 1648 - 1655
ISSN 0018-9456
DOI
Verlag New York, NY: IEEE
Freie Schlagworte S-parameters ; microwave measurements ; vector network analyzer (VNA) ; VNA calibration ; uncertainty of measurement ; sensitivity coefficients ; calibration standards
Zusammenfassung A novel method for the calibration of a vector network analyzer (VNA) is presented where a through connection of the VNA test ports, two known high-reflective calibration elements serving as calibration standards, and a termination of arbitrary unknown reflection are required. The latter has to be connected to both ports and may, for example, be a low-reflective load or a flush short. If it is possible to connect the same known high-reflective element to both ports, the VNA is then calibrated by the use of only one known standard. As only one two-port connection, namely a through, is required, the new method is advantageous especially for measurements in waveguides at very high frequencies if we compare it with the well-known through-reflect-line (TRL) and line-reflect-line (LRL) calibration methods where two two-port connections, namely a through and a line (TRL) or two lines (LRL) are needed. Measurements of various low- and high-reflective devices under test (DUTs) in the  WR-10 and WR-03 waveguide bands agreed well with TRL and LRL reference measurements, respectively. The influence of nonideal calibration elements or of reflection variations of the unknown termination on the S-parameter measurements of DUTs has also been theoretically and experimentally investigated.

Zitierung

Stumper, U. & Schrader, T. (2014). Calibration method for vector network analyzers using one or two known reflection standards. IEEE Transactions on Instrumentation and Measurement, 63(6), 1648–1655. https://doi.org/10.1109/tim.2013.2292754

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag