Several simple methods to test the humidity influence on the sensitivity of torque transducers are examined. Their suitability for use in calibration laboratories and important influences are discussed. The equivalence of former determinations of the humidity effect with extensive methods to the simplified methods is demonstrated.
Paper identifier: IMEKO-ACTA-03 (2014)-02-09
Kostenfreier Zugang
Open Access Gold
Rechteinformation
CC BY 3.0 ; Creative Commons Attribution 3.0 License
Zitierung
Brüge, A. (2014). Simplified measurements of the humidity coefficient of torque transducers in calibration laboratories. Acta IMEKO, 3(2), 32–38.