| Titel | Characterising a Si(Li) detector element for the SIXA X-ray spectrometer |
| Autor(in); Institution |
Tikkanen, T.; Observatory and Astrophysics Laboratory, University of Helsinki, FINLAND
Kraft, Stefan; 7.12, Röntgenradiometrie, PTB-Berlin
Scholze, Frank; 7.12, Röntgenradiometrie, PTB-Berlin
Thornagel, Reiner; 7.12, Röntgenradiometrie, PTB-Berlin
Ulm, Gerhard; 7.1, Photonenradiometrie, PTB-Berlin
|
| Quelle/Jahr | Nuclear Instruments and Methods in Physics Research A: 390 (1997), 3, 329 - 335 |
| ISSN | 0168-9002 |
| DOI | |
| Verlag | Amsterdam: North-Holland |
| Klassifikationscode | PACS: 95.55.Ka ; PACS: 07.85.Nc ; PACS: 29.40.Wk ; PACS: 85.30.De |
| Freie Schlagworte | Si(Li) detectors ; X-ray spectrometers ; Detector calibration ; X-ray response ; Spectral line shape |
| Zusammenfassung | The detection efficiency and response function of a Si(Li) detector element for the SIXA spectrometer have been determined in the 500 eV to 5 keV energy range using synchrotron radiation emitted at a bending magnet of the electron storage ring BESSY, which is a primary radiation standard. The agreement between the measured spectrum and the model calculation is better than 2%. |