| Zusammenfassung | For the first time, a EUROMET comparison was performed in the field of small DC currents below 1 nA. The capabilities of thirteen European and non-European NMIs for traceable calibrations of picoammeters have been compared. For that purpose, two different commercial picoammeters were used as travelling instruments. They were calibrated at current values of ±100 fA, ±1 pA, ±10 pA and ±100 pA. The accuracy achieved was to a large extent limited by the stability of the travelling instruments. The expanded relative uncertainty (k=2) of the reference values varied between 2.7·10-4 and 1.9·10-5, depending on the current ranges and the travelling instruments. There was a good agreement between the results of most of the participants. |