| Zugriffsnummer | 29017 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | Design of a four-crystal monochromator beamline for radiometry at BESSY II |
| Autor(in); Institution |
Krumrey, Michael; 7.12, Röntgenradiometrie, PTB-Berlin
|
| Quelle/Jahr | Journal of Synchrotron Radiation: 5 (1998), 1, 6 - 9 |
| ISSN | 0909-0495 (PRINT) ; 1600-5775 (ONLINE) |
| DOI | |
| Verlag | Oxford: Blackwell |
| Freie Schlagworte | four-crystal monochromator ; radiometry ; X-ray optics |
| Zusammenfassung | A four-crystal monochromator beamline will be part of the radiometry laboratory that the Physikalisch-Technische Bundesanstalt will install at the new storage ring BESSY II. The most important design criteria for the beamline are the tunability of the photon energy in a wide spectral range from 1.75 to 10 keV, the high spectral purity of the radiation, as well as the good reproducibility of the absolute photon flux. |