Titel Design of a four-crystal monochromator beamline for radiometry at BESSY II
Autor(in); Institution
Krumrey, Michael; 7.12, Röntgenradiometrie, PTB-Berlin
Quelle/Jahr Journal of Synchrotron Radiation: 5 (1998), 1, 6 - 9
ISSN 0909-0495 (PRINT) ; 1600-5775 (ONLINE)
DOI
Verlag Oxford: Blackwell
Freie Schlagworte four-crystal monochromator ; radiometry ; X-ray optics
Zusammenfassung A four-crystal monochromator beamline will be part of the radiometry laboratory that the Physikalisch-Technische Bundesanstalt will install at the new storage ring BESSY II. The most important design criteria for the beamline are the tunability of the photon energy in a wide spectral range from 1.75 to 10 keV, the high spectral purity of the radiation, as well as the good reproducibility of the absolute photon flux.

Zitierung

Krumrey, M. (1998). Design of a four-crystal monochromator beamline for radiometry at BESSY II. Journal of Synchrotron Radiation, 5(1), 6–9. http://doi.org/10.1107/s0909049597011825