| Zugriffsnummer | 29015 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | Measurement of the electron energy and energy spread at the electron storage ring BESSY I |
| Autor(in); Institution |
Klein, Roman; 7.12, Röntgenradiometrie, PTB-Berlin
Mayer, T.; BESSY GmbH, Berlin, GERMANY
Kruske, P.; BESSY GmbH, Berlin, GERMANY
Thornagel, Reiner; 7.12, Röntgenradiometrie, PTB-Berlin
Ulm, Gerhard; 7.1, Photonenradiometrie, PTB-Berlin
|
| Quelle/Jahr | Journal of Synchrotron Radiation: 5 (1998), 3, 392 - 394 |
| ISSN | 0909-0495 |
| DOI | |
| Verlag | Oxford: Blackwell |
| Freie Schlagworte | metrology ; storage rings ; gamma-ray spectroscopy ; gamma-ray sources ; special relativity |
| Zusammenfassung | Knowledge of the electron energy with a small uncertainty is necessary for the Physikalisch-Technische Bundesanstalt (PTB) to operate the electron storage ring BESSY I, and the future BESSY II, as a primary radiation source standard of calculable synchrotron radiation. At BESSY I the electron energy can now be measured either by the long-established method of resonant spin depolarization or by the newly set up method of Compton backscattering (CBS) of laser photons (CO2 laser, λ = 10.6 μm). Results obtained at different electron energies by these two independent methods are presented. They agree within a relative uncertainty of better than 10-4. The advantages and disadvantages of these two complementary techniques are described and applications of CBS for the measurement of other storage-ring parameters, e.g. the electron energy spread, are given. |