| Zugriffsnummer | 28614 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | Technique for the calibration of thermal voltage converters using a Josephson waveform synthesizer and a transconductance amplifier |
| Autor(in); Institution |
Budovsky, Ilya; National Measurement Institute (NMI), Lindfield, NSW, AUSTRALIA
Behr, Ralf; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Palafox Gamir, Luis; 2.6, Elektrische Quantenmetrologie, PTB-Braunschweig
Djordjevic, Sophie; Laboratoire National de Métrologie et d’Essais (LNE), Trappes, FRANCE
Hagen, Thomas; National Measurement Institute (NMI), Lindfield, NSW, AUSTRALIA
|
| Quelle/Jahr | Measurement Science and Technology: 23 (2012), 12, 124005-1 - 124005-6 |
| ISSN | 0957-0233 |
| DOI | |
| URL | |
| Verlag | Bristol: IOP |
| Klassifikationscode | PACS: 06.20.fb ; PACS: 84.30 ; PACS: 06.20.Dk |
| Freie Schlagworte | Josephson array ; thermal converter ; ac–dc transfer difference ; signal synthesis ; voltage measurement |
| Zusammenfassung | This paper describes a simple and effective technique for calibrating thermal converters for an ac–dc voltage transfer difference using a Josephson waveform synthesizer and a transconductance amplifier. Preliminary measurements and uncertainty analysis confirm the possibility of achieving systematic uncertainties below 0.1 µV V−1 at frequencies up to 1 kHz. |
Zitierung
Budovsky, I., Behr, R., Palafox Gamir, L., Djordjevic, S., & Hagen, T. (2012). Technique for the calibration of thermal voltage converters using a Josephson waveform synthesizer and a transconductance amplifier. Measurement Science and Technology, 23(12), 124005-1–124005-6. https://doi.org/10.1088/0957-0233/23/12/124005