| Zugriffsnummer | 28558 |
| Dokumenttyp | Zeitschriftenartikel |
| Peer Review | unbekannt |
| Sprache | Englisch |
| Titel | EURAMET.RI(I)-S7 comparison of alanine dosimetry systems for absorbed dose to water measurements in gamma- and x-radiation at radiotherapy levels |
| Autor(in); Institution |
Garcia, Tristan; CEA, LIST, Laboratoire National Henri Becquerel (LNE-LNHB), Gif-sur-Yvette CEDEX, FRANCE
Anton, Mathias; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig
Sharpe, Peter; National Physical Laboratory (NPL), Teddington, UK
|
| Quelle/Jahr | Metrologia: 49 (2012), Tech. Suppl., 18 S. |
| Artikelnummer | 06004 |
| ISSN | 0026-1394 |
| DOI | |
| Verlag | Bristol: IOP |
| Freie Schlagworte | dosimetry ; cobalt and MVX beams ; alanine ; Intercomparison ; Euramet |
| Zusammenfassung | An intercomparison of the alanine dosimetry systems of LNHB, PTB and NPL was carried out. Each institution irradiated probes for the two other ones as well in Co-60 as in a selected MVX reference field. Results of the dose evaluation were collected by BIPM before distribution to the authors. The results of the three laboratories are consistent, the largest deviation between stated and measured dose that occurred was 1%. |
Zitierung
Garcia, T., Anton, M., & Sharpe, P. (2012). EURAMET.RI(I)-S7 comparison of alanine dosimetry systems for absorbed dose to water measurements in gamma- and x-radiation at radiotherapy levels. Metrologia, 49(Tech. Suppl.), 18 S. https://doi.org/10.1088/0026-1394/49/1a/06004