Titel EURAMET.RI(I)-S7 comparison of alanine dosimetry systems for absorbed dose to water measurements in gamma- and x-radiation at radiotherapy levels
Autor(in); Institution
Garcia, Tristan; CEA, LIST, Laboratoire National Henri Becquerel (LNE-LNHB), Gif-sur-Yvette CEDEX, FRANCE
Anton, Mathias; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig
Sharpe, Peter; National Physical Laboratory (NPL), Teddington, UK
Quelle/Jahr Metrologia: 49 (2012), Tech. Suppl., 18 S.
Artikelnummer 06004
ISSN 0026-1394
DOI
Verlag Bristol: IOP
Freie Schlagworte dosimetry ; cobalt and MVX beams ; alanine ; Intercomparison ; Euramet
Zusammenfassung An intercomparison of the alanine dosimetry systems of LNHB, PTB and NPL was carried out. Each institution irradiated probes for the two other ones as well in Co-60 as in a selected  MVX reference field. Results of the dose evaluation were collected by BIPM before distribution to the authors. The results of the three laboratories are consistent, the largest deviation between stated and measured dose that occurred was 1%.