| Zugriffsnummer | 28265 |
| Dokumenttyp | Konferenzartikel |
| Sprache | Englisch |
| Titel | Inter-laboratory comparison of reference torque wrench calibration between NMIJ and PTB in the range of from 200 N·m To 2 kN·M |
| Autor(in); Institution |
Ogushi, Koji; National Metrology Institute of Japan, AIST, JAPAN
Nishino, Atsuhiro; National Metrology Institute of Japan, AIST, JAPAN
Maeda, Kojoi; National Metrology Institute of Japan, AIST, JAPAN
Ueda, Kazunaga; National Metrology Institute of Japan, AIST, JAPAN
Brüge, Andreas; 1.2, Festkörpermechanik, PTB-Braunschweig
Röske, Dirk; 1.2, Festkörpermechanik, PTB-Braunschweig
|
| Quelle/Jahr | Proceedings of XX IMEKO World Congress : Metrology for Green Growth:(2012), 6 S. |
| Availability | [USB-Stick] ; file name: 413-F-O-TC3-116-77.pdf |
| ISBN | 978-89-9500005-2 |
| URL | |
| Verlag | Busan: |
| Konferenzangaben | XX IMEKO World Congress: Metrology for Green Growth, Busan, 09-14, September, 2012, Republic of Korea |
| Freie Schlagworte | torque ; Torque wrench ; Reference torque wrench ; Comparison ; Square drive |
| Zusammenfassung | Inter-laboratory comparison of reference torque wrench calibration was conducted between NMIJ and PTB using a torque transducer in the form of a torque wrench, "TTS/3kNm," as a transfer device. Although the rated capacity of the transducer is 3 kN·m, the calibration range of from 200 N·m to 2 kN·m was selected. Each loading time schedule used in daily calibration was also adopted in this comparison. Calibration results obtained by both laboratories coincided with the range of uncertainties. |
Zitierung
Ogushi, K., Nishino, A., Maeda, K., Ueda, K., Brüge, A., & Röske, D. (2012). Inter-laboratory comparison of reference torque wrench calibration between NMIJ and PTB in the range of from 200 N·m To 2 kN·M. XX IMEKO World Congress: Metrology for Green Growth, Busan, 09-14, September, 2012, Republic of Korea.