| Zugriffsnummer | 27852 |
| Dokumenttyp | Konferenzartikel in Zeitschrift |
| Sprache | Englisch |
| Titel | Recent developments in radon metrology: new aspects in the calibration of radon, thoron and progeny devices |
| Autor(in); Institution |
Röttger, Annette; 6.1, Radioaktivität, PTB-Braunschweig
Honig, Anja; 6.1, Radioaktivität, PTB-Braunschweig
|
| Quelle/Jahr | 6th Conference on Protection against Radon at home and at work. Radiation Protection Dosimetry: 145 (2011), 2/3, 260 - 266 |
| ISSN | 0144-8420 |
| DOI | |
| Verlag | Oxford: Oxford Univ. Press |
| Konferenzangaben | 6th Conference on Protection against Radon at Home and at Work, Prague, 13-17, September, 2009, Czech Republic |
| Zusammenfassung | Due to the importance of reliable measurements of radon activity concentration, one of the past developments in metrology was applied to the field of radon, thus meeting two basic needs: (1) the harmonisation of metrology within the scope of the mutual recognition arrangement, an arrangement drawn up by the International Committee of Weights and Measures for the mutual recognition of national standards and of calibrations issued by national metrology institutes and (2) the increased demands of the European Atomic Energy Community (EURATOM) directive, transferred into national radiation protection regulations with regard to natural radioactivity and its quality-assured measurements. This paper gives an overview of typical technical procedures in the radon-measuring technique group of PTB, covering all aspects of reference atmospheres (primary standards) for radon, thoron and their respective progenies. |