Zugriffsnummer 27622
Dokumenttyp Zeitschriftenartikel
Peer Review unbekannt
Sprache Englisch
Titel How the cross-sectional discontinuity between ear canal and probe affects the ear canal length estimation
Autor(in); Institution
Zebian, Makram; 1.6, Schall, PTB-Braunschweig
Hensel, Johannes; 1.6, Schall, PTB-Braunschweig
Fedtke, Thomas; 1.6, Schall, PTB-Braunschweig
Quelle/Jahr Journal of the Acoustical Society of America: 132 (2012), 1, EL8 - EL14
ISSN 0001-4966 (PRINT) ; 1520-8524 (ONLINE)
DOI
URL
Verlag Melville, NY: AIP
Klassifikationscode PACS: 43.20.Ks ; PACS: 43.30.Ma
Freie Schlagworte finite element analysis ; otoacoustic emissions
Zusammenfassung Many ear canal probes both deliver and measure sound via narrow tubes. This study investigates the effect of the cross-sectional discontinuity at the interface between ear canal models and the connecting tubes of a commercially available otoacoustic emission probe on the "acoustically" estimated cavity lengths. Rigid cavities having the same length but different diameters were produced, and modeled by the finite element method. Cavities with a diameter larger than 8 mm had acoustic lengths that considerably overestimated the real geometry. A length correction was derived, which, in most applications, compensates for the measurement errors emerging from the discontinuity effects

Zitierung

Zebian, M., Hensel, J., & Fedtke, T. (2012). How the cross-sectional discontinuity between ear canal and probe affects the ear canal length estimation. Journal of the Acoustical Society of America, 132(1), EL8–EL14. https://doi.org/10.1121/1.4723565

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag