| Zugriffsnummer | 27511 |
| Dokumenttyp | Konferenzartikel in Zeitschrift |
| Peer Review | unbekannt |
| Sprache | Englisch |
| Titel | Novel reference field for pulsed photon radiation for research and type testing |
| Autor(in); Institution |
Klammer, Jana; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Hupe, Oliver; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Roth, Jürgen; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
|
| Quelle/Jahr | Radiation Protection Dosimetry: 151 (2012), 3, 478 - 482 |
| ISSN | 0144-8420 |
| DOI | |
| Verlag | Oxford: Oxford University Press |
| Konferenzangaben | 13th IRPA, International Congress of the International Radiation Protection Association, Glasgow, 13-18, May, 2012, Scotland |
| Zusammenfassung | Currently, ca. 70 % of the occupationally exposed persons in Germany are working in pulsed radiation fields, mainly in the medical sector. It has been known for a few years that active electronic dosemeters exhibit considerable deficits or can even fail completely in pulsed fields. Type test requirements for dosemeters exist only for continuous radiation. Owing to the need of a reference field for pulsed photon radiation and accordingly to the upcoming type test requirements for dosemeters in pulsed radiation, the Physikalisch-Technische Bundesanstalt has developed a novel X-ray reference field for pulsed Photon radiation in cooperation with a manufacturer. this reference fiekd, geared to the main application in the field of medicine, has been well characterised and is now available for research and tpye testing of dosemeters in pulsed photon radiation. |