Titel Design, evaluation and application of a sixteen channel transmit/receive surface coil array for cardiac MRI at 7T [poster]
Autor(in); Institution
Thalhammer, C.; Berlin Ultra-High Field Facility (BUFF), Max Delbrück Center for Molecular Medicine (MDC), Berlin, GERMANY
Renz, W.; Berlin Ultra-High Field Facility (BUFF), Max Delbrück Center for Molecular Medicine (MDC), Berlin, GERMANY
Pfeiffer, Harald; 8.1, Medizinische Messtechnik, PTB-Berlin
Rieger, J.; Berlin Ultra-High Field Facility (BUFF), Max Delbrück Center for Molecular Medicine (MDC), Berlin, GERMANY
Winter, L.; Berlin Ultra-High Field Facility (BUFF), Max Delbrück Center for Molecular Medicine (MDC), Berlin, GERMANY
Hezel, F.; Berlin Ultra-High Field Facility (BUFF), Max Delbrück Center for Molecular Medicine (MDC), Berlin, GERMANY
Seifert, Frank; 8.1, Medizinische Messtechnik, PTB-Berlin
Hoffmann, Werner; 8.1, Medizinische Messtechnik, PTB-Berlin
Seemann, Rainer; 8.1, Medizinische Messtechnik, PTB-Berlin
Niendorf, T.; Berlin Ultra-High Field Facility (BUFF), Max Delbrück Center for Molecular Medicine (MDC), Berlin, GERMANY
Quelle/Jahr Proceedings of the International Society for Magnetic Resonance in Medicine ; ISMRM 19th annual meeting & exhibition: 19 (2011), 326
Availability [CD-ROM] ; file name: 326.pdf
ISSN 1545-4428
Verlag Berkeley, Calif.: ISMRM
Konferenzangaben ISMRM-ESMRMB Joint Annual Meeting ; ISMRM 19th Scientific Meeting, Montreal, 07-13, May, 2011, Kanada
Zusammenfassung Introduction: Detrimental effects bear the potential to spoil the signal-to-noise (SNR) and contrast-to-noise (CNR) benefits of cardiac MR (CMR) at ultrahigh (≥7T) fields. B1+-inhomogeneities and signal voids represent the main challenges. Various coil concepts have been proposed to tackle these issues, enabling CMR images at 7 T. Modulating B1+ in two directions has been demonstrated to enhance the RF efficiency of a coil array for head imaging. Also, parallel imaging benefits from two-dimensional acceleration. For all these reasons, this work proposes a two dimensional 16-channel transceive array using loop elements and demonstrates its applicability for enhanced B1+ homogeneity and improved parallel imaging performance for cardiac MRI at 7 T. Methods: The coil array consists of a planar posterior section (Fig. 1a) and a bent anterior section (Fig. 1b). Both sections comprise 8 rectangular elements (2 x 4 array) with an effective size of 13 cm x 6 cm each (Fig. 1a). The conductors have a width of 2 cm. Adjacent elements have a common conductor with an integrated, adjustable capacitor for decoupling of neighboring elements. To reduce the radiation losses of the array, an RF shield, made of patches of slotted copper foil, was placed at a distance of 2 cm behind the conductors. Unbalanced currents on the coaxial cables were minimized by cable traps. All 16 elements were connected to multipurpose transmit/receive switch boxes with integrated low-noise preamplifiers. Electro-magnetic (EM) field and SAR simulations were performed using CST Studio Suite 2010 (CST AG, Darmstadt, Germany) together with voxel models from the Virtual Family. The RF characteristics were measured using an 8-channel network analyzer (Rohde & Schwarz, Munich, Germany). The MR experiments were conducted on a 7 T scanner (Siemens Healthcare, Erlangen, Germany), equipped with an RF amplifier providing a peak power of 8 kW. The amplifier output was split into 16 equal-intensity signals with relative phase offsets of a multiple of 2π/16 by means of a home-built 16 × 16 Butler matrix. These phases modes were slightly modified by phase-shifting coaxial cables to match them to the coil's geometry. Cardiac MR was performed in 6 subjects using single breath-hold 2D CINE FLASH sequences in conjunction with retrospective acoustic cardiac gating (MRI Tools GmbH, Berlin, Germany) and GRAPPA based parallel imaging. Results: The RF characteristics were below -15 dB for element-coupling, -20 dB or less for reflection and an average of 5 for QU/QL. The values were measured on different volunteers without subject-specific tuning and matching. Noise correlation of the array was below 0.3 between all elements (Fig. 1c). SAR values, derived from the EM simulations, were well below permitted limits of the IEC guidelines for an average power of 30 W over 6 minutes (Fig. 1d) (including the application of different phase configurations). A set of 3 fixed phase settings was used to avoid RF shading in the cardiac region for all subjects investigated. The acquired images exhibit a rather uniform intensity over a large region with a high myocardium/blood contrast as demonstrated in Fig. 2 for various long axis and short axis standard cardiac views. The overall image quality enabled the visualization of subtle anatomic structures such as pericardium, mitral and tricuspid valves and their associated papillary muscles and trabeculae. Accelerated 2D CINE FLASH provided excellent image quality even for R=4 as demonstrated in Fig. 3. Discussion: Our results demonstrate that the proposed two-dimensional 16 channel coil array based on loop design is capable of producing highly uniform cardiac/torso images at 7 T. Combining a large number of surface coil elements yielded an excellent SNR and CNR and facilitated the depiction of subtle anatomical cardiac details. The excellent parallel imaging performance enabled the acquisition of high resolution images with satisfactory SNR within one breath-hold (Fig. 2d). We anticipate further improvement in image quality by using more advanced B1+ shimming.

Zitierung

Thalhammer, C., Renz, W., Pfeiffer, H., Rieger, J., Winter, L., Hezel, F., Seifert, F., Hoffmann, W., Seemann, R., & Niendorf, T. (2011). Design, evaluation and application of a sixteen channel transmit/receive surface coil array for cardiac MRI at 7T [poster]. ISMRM-ESMRMB Joint Annual Meeting ; ISMRM 19th Scientific Meeting, Montreal, 07-13, May, 2011, Kanada.