Zugriffsnummer 26314
Dokumenttyp Zeitschriftenartikel
Peer Review unbekannt
Sprache Englisch
Titel Synchrotron radiation-based bilateral intercomparison of ultraviolet source calibrations
Autor(in); Institution
Arp, U.; Electron and Optical Physics Division, National Institute of Standards and Technology, Gaithersburg, USA
Klein, Roman; 7.2, Radiometrie mit Synchrotronstrahlung, PTB-Berlin
Li, Z.; Electron and Optical Physics Division, National Institute of Standards and Technology, Gaithersburg, USA
Paustian, Wolfgang; 7.2, Radiometrie mit Synchrotronstrahlung, PTB-Berlin
Richter, Mathias; 7.1, Röntgenmesstechnik mit Synchrotronstrahlung, PTB-Berlin
Shaw, P.-S.; Electron and Optical Physics Division, National Institute of Standards and Technology, Gaithersburg, USA
Thornagel, Reiner; 7.2, Radiometrie mit Synchrotronstrahlung, PTB-Berlin
Quelle/Jahr Metrologia: 48 (2011), 5, 261 - 267
ISSN 0026-1394
DOI
Verlag Bristol: Inst. of Physics Publ.
Klassifikationscode PACS: 06.20.fb
Freie Schlagworte Instrumentation and measurement
Zusammenfassung We report on a successful bilateral intercomparison between the Physikalisch-Technische Bundesanstalt (PTB) and the National Institute of Standards and Technology (NIST). In both laboratories deuterium lamps were calibrated using the calculability of synchrotron radiation. However, the methods applied in the two laboratories differed significantly. Whereas the calibrations at PTB were performed in vacuum and led to spectral radiant intensities, the measurements at NIST were done in air and resulted in spectral irradiances. Therefore, for this bilateral comparison we not only proved the equivalence of the two measurement procedures, but also the validity of the conversion between the different spectro-radiometric quantities. This successful comparison re-establishes the usefulness of deuterium lamps as transfer standards in the ultraviolet spectral range.

Zitierung

Arp, U., Klein, R., Li, Z., Paustian, W., Richter, M., Shaw, P.-S., & Thornagel, R. (2011). Synchrotron radiation-based bilateral intercomparison of ultraviolet source calibrations. Metrologia, 48(5), 261–267. https://doi.org/10.1088/0026-1394/48/5/004

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag