| Zugriffsnummer | 25263 |
| Dokumenttyp | Zeitschriftenartikel |
| Peer Review | unbekannt |
| Sprache | Englisch |
| Titel | Comparison of the radiation temperatur scales between MIKES and PTB |
| Autor(in); Institution |
Ojanen, Maija; Helsinki University of Technology (TKK), Metrology Research Institute, Espoo, FINLAND
Anhalt, Klaus; 7.3, Detektorradiometrie und Strahlungsthermometrie, PTB-Berlin
Hartmann, Jürgen; 7.3, Detektorradiometrie und Strahlungsthermometrie, PTB-Berlin
Schiller, Stephan; 7.3, Detektorradiometrie und Strahlungsthermometrie, PTB-Berlin
Weckström, T.; Centre for Metrology and Accreditation (MIKES), Espoo, FINLAND
Kärhä, P.; Helsinki University of Technology (TKK), Metrology Research Institute, Espoo, FINLAND
Heinonen, M.; Centre for Metrology and Accreditation (MIKES), Espoo, FINLAND
Ikonen, Erkki; Helsinki University of Technology (TKK), Metrology Research Institute, Espoo, FINLAND; Centre for Metrology and Accreditation (MIKES), Espoo, FINLAND
|
| Quelle/Jahr | Measurement: 43 (2010), 2, 183 - 189 |
| ISSN | 0263-2241 |
| DOI | |
| Verlag | Amsterdam [u.a.]: Elsevier |
| Freie Schlagworte | Filter radiometer ; Intercomparison ; Linear radiation thermometer ; Radiation temperature scale |
| Zusammenfassung | The radiation temperature scales of PTB and MIKES were compared in the range of 1570–2770 K using four filter radiometers of MIKES, one filter radiometer of PTB, and linear radiation thermometers of both MIKES and PTB. The agreement was partial: two filter radiometers and the linear radiation thermometer of MIKES agreed well with the equipment of PTB, while two filter radiometers deviated from the other equipment. To get a deeper understanding of the reasons of the deviation, the results were studied in terms of both temperature and radiance. The deviation was found constant in terms of radiance. A correction factor was calculated for the radiance level with help of a temperature fixed-point calibration. The use of the correction factor improved the results. |