Zugriffsnummer 24454
Dokumenttyp Zeitschriftenartikel
Peer Review unbekannt
Sprache Englisch
Titel Influence of the uniformity of the slab phontom illumination on the calibration of personal dosemeters
Autor(in); Institution
Ankerhold, Ulrike; 6.2, Dosimetrie für Strahlentherapie und Röntgendiagnostik, PTB-Braunschweig
Ambrosi, Peter; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Quelle/Jahr Radiation Protection Dosimetry: 140 (2010), 1, 9 - 15
ISSN 0144-8420
DOI
Verlag Oxford: Oxford University Press
Freie Schlagworte calibration ; personal dosemeters ; Hp(10) ; ISO water slab phontom ; irradiation geometry
Zusammenfassung For the calibration of personal dosemeters in terms of Hp(10), the ISO water slab phantom shall be used. The irradiation geometry must be chosen such that the field diameter at the phantom front surface is large enough to irradiate the phantom completely and, in addition, homogeneously. However, in practice it is not always possible to achieve this, especially if high dose rates are required and hence the irradiation must be performed at short distances. The subject of this investigation is to determine the influence of the beam diameter on the calibration result or, more correctly, on the quantity value of Hp(10). Measurements were performed with a PTB-developed secondary standard chamber for Hp(10) at several radiation qualities of the ISO narrow-spectrum series. The results show that a reduced beam diameter of only 250 mm decreases the quantity value of Hp(10) by a maximum of 5 %. This can be considered by adding an additional standard uncertainty of a maximum of 5 % for the quantity value of Hp(10).

Zitierung

Ankerhold, U. & Ambrosi, P. (2010). Influence of the uniformity of the slab phontom illumination on the calibration of personal dosemeters. Radiation Protection Dosimetry, 140(1), 9–15. https://doi.org/10.1093/rpd/ncq036

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag