Zugriffsnummer 22993
Dokumenttyp Konferenzartikel in Zeitschrift
Sprache Englisch
Titel Influence of nonideal calibration items on S-parameter uncertainties applying the SOLR calibration method
Autor(in); Institution
Stumper, Ulrich; PTB-Braunschweig, retired staff
Quelle/Jahr IEEE Transactions on Instrumentation and Measurement: 58 (2009), 4, 1158 - 1163
ISSN 0018-9456
DOI
Verlag New York, NY: IEEE
Konferenzangaben Conference on Precision Electromagnetic Measurements (CPEM), Boulder, Colo., 08-13, June, 2008, USA
Freie Schlagworte measuring uncertainty ; scattering parameter measurement ; sensitivity coefficients ; SOLR calibration ; SOLT calibration ; vector network analyzer
Zusammenfassung For the seven-term, short-open-load-reciprocal (SOLR) calibration method to calibrate a four-sampler vector network analyzer (VNA), the expressions for the deviations of the seven error terms and of the measured S-parameters of two-port test objects from their true values are presented; these deviations are caused by the deviations of the reflections of the calibration elements (shorts, opens, and loads) from their ideal values. For the SOLR calibration, a reciprocal unknown network is used instead of a known through. The influence of a slightly nonreciprocal network on both the error term and the S-parameter deviations is also considered. The obtained sensitivity coefficients are useful for establishing the type-B uncertainty budget for S-parameter measurements. Moreover, the results give valuable information for the design of the calibration elements.

Zitierung

Stumper, U. (2009). Influence of nonideal calibration items on S-parameter uncertainties applying the SOLR calibration method. IEEE Transactions on Instrumentation and Measurement, 58(4), 1158–1163. https://doi.org/10.1109/tim.2008.2006962

Exportieren