Titel Deficiencies of active electronic radiation protection dosemeters in pulsed fields
Autor(in); Institution
Ankerhold, Ulrike; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Hupe, Oliver; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Ambrosi, Peter; 6.3, Strahlenschutzdosimetrie, PTB-Braunschweig
Quelle/Jahr Radiation Protection Dosimetry: 135 (2009), 3, 149 - 153
ISSN 0144-8420
DOI
Verlag Oxford: Oxford University Press
Zusammenfassung Nowadays nearly all radiation fields used for X-ray diagnostics are pulsed. These fields are characterised by a high dose rate during the pulse and a short pulse duration in the range of a few milliseconds. The use of active electronic dosemeters has increased in the past few years, but these types of dosemeter might possibly not measure reliably in pulsed radiation fields. Not only personal dosemeters but also area dosemeters which are sued mainly for dose rate measurements are concerned. These cannot be substituted by using passive dosemeter types. The characteristics of active electronic dosemeters determined in a continuous radiation field cannot be transferred to those in pulsed fields. Some provisional measurements with typical electronic dosemeters in pulsed radiation fields are presented to reveal this basic problem.