| Zugriffsnummer | 22705 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | Monte Carlo simulation of Auger-electron spectra |
| Autor(in); Institution |
Grau Carles, A.; Instituto de Matematicas y Fisica Fundamental, Madrid, SPAIN
Kossert, Karsten; 6.1, Radioaktivität, PTB-Braunschweig
|
| Quelle/Jahr | Applied Radiation and Isotopes: 67 (2009), 1, 192 - 196 |
| ISSN | 0969-8043 |
| DOI | |
| Verlag | Amsterdam: Elsevier |
| Freie Schlagworte | Auger-electron spectra ; 125I, 123I, 111In ; Liquid scintillation counting |
| Zusammenfassung | A procedure to calculate the complex spectra of electron-capture nuclides which simultaneously eject several electrons and X-rays with different energies is presented. The model is applied to compute spectra of the radionuclides 125I, 123I and 111In. The spectra are then compared with experimental spectra obtained by means of liquid scintillation counting. To this end, the computed spectra were transformed to allow for the nonlinear response function for a liquid scintillator, chemical quenching, as well as the Wallac-type amplifier used for the measurements. The calculated spectra are important for applications of free parameter models in liquid scintillation counting and also for studying the impact of electron-capture nuclides on DNA. |