Zugriffsnummer 22296
Dokumenttyp Konferenzartikel
Sprache Englisch
Titel Position and time sensitive photon counting detector with image charge delay-line readout
Autor(in); Institution
Czasch, Achim; University Frankfurt, Frankfurt, GERMANY; RoentDek GmbH, GERMANY
Dangendorf, Volker; 6.5, Neutronenstrahlung, PTB-Braunschweig
Milnes, Jamers; Photek Ltd., St. Leonards on Sea, East Sussex, UK
Schössler, Sven; University Frankfurt, Frankfurt, GERMANY
Lauck, Ronald; 6.5, Neutronenstrahlung, PTB-Braunschweig
Spillmann, Uwe; RoentDek GmbH, GERMANY
Howorth, Jon; Photek, Ltd., UK
Jagutzki, Ottmar; University Frankfurt, Frankfurt, GERMANY; RoentDek GmbH, GERMANY
Quelle/Jahr Advanced Photon Counting Techniques II:(2007), 67710W-1 - 67710W-12
Schriftenreihe Proceedings of SPIE: 6771
Herausgeber(in)
Becker, Wolfgang
ISBN 978-0-8194-6931-1
DOI
Verlag Bellingham, Wash.: SPIE
Konferenzangaben Optics East, Boston, 09-12, September, 2007, USA
Freie Schlagworte photo-multiplier tube ; MCP ; image intensifier ; delay-line ; photon counting ; 3d imaging ; X-ray polarimetry ; fast-neutron radiography ; homeland security
Zusammenfassung We have developed single photon counting image intensifier tubes combining position and time information readout with at least 500 × 500 pixels and sub-nanosecond time resolution. This image intensifier type uses a resistive screen instead of a phosphor screen and the image charge pickup anode is placed outside the sealed tube. We present a novel delay-line anode design which allows for instance detecting simultaneously arriving pairs of photons. Due to the very low background this technique is suited for applications with very low light intensity and especially if a precise time tagging for each photon is required. We show results obtained with several anode types on a 25 mm image intensifier tube and a 40 mm open-face MCP detector and discuss the performance in neutron radiography, e.g. for homeland security, and the prospects for applications like Fluorescence Life-time Imaging Microscopy (FLIM), astronomy and Xray polarimetry.

Zitierung

Czasch, A., Dangendorf, V., Milnes, J., Schössler, S., Lauck, R., Spillmann, U., Howorth, J., & Jagutzki, O. (2007). Position and time sensitive photon counting detector with image charge delay-line readout. Optics East, Boston, 09-12, September, 2007, USA. https://doi.org/10.1117/12.736935