| Titel | Uncertainties of VNA S-parameter measurements applying the short-open-load-reciprocal (SOLR) calibration method |
| Autor(in); Institution |
Stumper, Ulrich; 2.2, Hochfrequenz und Felder, PTB-Braunschweig; PTB-Braunschweig, retired staff
|
| Quelle/Jahr | 2008 Conference on Precision Electromagnetic Measurements digest:(2008), 438 - 439 |
| Herausgeber(in) |
Cookson, Alan H.
|
| ISBN | 978-1-424-42399-6 (print) ; 978-1-424-42400-9 (CD-ROM) |
| Verlag | Piscataway, NJ: IEEE |
| Konferenzangaben | 2008 Conference on Precision Electromagnetic Measurements, Broomfield, Colorado, 08-13, June, 2008, USA |
| Zusammenfassung | For the 7-term SOLR calibration method to calibrate a four-sampler vector network analyser (VNA), the deviations of the seven error terms, as well as of the the measured S-parameters are determined due to the deviations of the reflections of the calibration elements (Short, Open and Load) from their true values. For the SOLR calibration, a reciprocal unknown network is used instead of a known Through. The influence of a slightly non-reciprocal network on error term and S-parameter deviations is also considered. |