Titel Final report on the CCPR-S2 supplementary comparison of area measurements of apertures for radiometry
Autor(in); Institution
Litorja, Maritoni; National Institute of Standards and Technology (NIST), Gaithersburg, USA
Fowler, Joel; National Institute of Standards and Technology (NIST), Gaithersburg, USA
Hartmann, Jürgen; 7.3, Hochtemperatur- und Vakuumphysik, PTB-Berlin
Fox, Nigel; National Physical Laboratory (NPL), Teddington, UK
Stock, Michael; Bureau International des Poids et Mesures (BIPM), Sèvres, FRANCE
Razet, Annick; Bureau National des Metrologie, Institut National de Metrologie (BNM-INM/CNAM), Paris, FRANCE
Khlevnoy, Boris; All-Russian Research Institute for Optical and Physical Measurements (VNIIOFI), Moscow, RUSSIA
Ikonen, Erkki; Centre for Metrology and Accreditation (CMA), Mittatekniikan Keskus (MIKES), Helsinki, FINLAND
Machacs, M.; National Office of Measures (OMH), Budapest, HUNGARY
Doytchinov, Kostadin; National Research Council of Canada (NRC), Ottawa, CANADA
Quelle/Jahr Metrologia: 44 (2007), Tech. Suppl.
Artikelnummer 02002
Availability [online only]
ISSN 0026-1394 ; 1681-7575
DOI
URL
Verlag Sèvres: BIPM
Zusammenfassung The accuracy of aperture area measurements is vital to many radiometric and photometric measurements. The Consultative Committee on Photometry and Radiometry (CCPR) undertook an international comparison of the methods by which member laboratories measure the areas of apertures used for radiometry. Nine laboratories measured eight different apertures varying in diameter size, fabrication method, material and edge type. This paper includes a description of the apertures used for the comparison, the results of the comparison categorized by artifact and by laboratory, and a brief discussion of the method used by each laboratory along with its uncertainty budget. Orig.-Titel des Reports: Report on the CCPR-S2: Supplementary Comparison of Area Measurements of Apertures for Radiometry

Zitierung

Litorja, M., Fowler, J., Hartmann, J., Fox, N., Stock, M., Razet, A., Khlevnoy, B., Ikonen, E., Machacs, M., & Doytchinov, K. (2007). Final report on the CCPR-S2 supplementary comparison of area measurements of apertures for radiometry. Metrologia, 44(Tech. Suppl.). https://doi.org/10.1088/0026-1394/44/1a/02002