Zugriffsnummer 20366
Dokumenttyp Konferenzartikel in Zeitschrift
Peer Review unbekannt
Sprache Englisch
Titel Intercomparison of copper fixed-point cells by using Pt/Pd thermocouples [poster]
Autor(in); Institution
Edler, Frank; 7.4, Temperatur, PTB-Berlin
Anagnostou, M.; Hellenic Institute of Metrology (EIM), Thessaloniki, GREECE
Bojkovski, Jovan; University of Ljubljana, Faculty of Electrical Engineering, Laboratory of Metrology and Quality (MIRS/FE-LMK) Ljubljana, SLOVENIA
Gaita, S.; National Institute of Metrology (INM), Bucurest, ROMANIA
Garcia, C.; Centro Espanol de Metrologia (CEM), Madrid, SPAIN
Grudniewicz, E.; Central Office of Measures (GUM), Warsaw, POLAND
Helgesen, F.; Norwegian Metrology Service (JV), NORWAY
Ivarsson, J.; Swedisch National Testing and Research Institute (SP), Boras, SWEDEN
Pauza, A.; Semiconductor Physics Institute (VMT/PFI), LITHUANIA
Rosenkranz, P.; Bundesamt für Eich- und Vermessungswesen (BEV), Wien, AUSTRIA
Smid, M.; Czech Metrology Institute (CMI), Praha, CZECH REPUBLIC
Weckström, Thua; Centre for Metrology and Accreditation (MIKES), Helsinki, FINLAND
Zvizdic, Davor; Laboratory for Process Measurement (LPM), CROATIA
Quelle/Jahr International Journal of Thermophysics: 29 (2008), 1, 171 - 180
ISSN 0195-928X (PRINT) ; 1572-9567 (ONLINE)
DOI
Verlag Dordrecht: Springer
Konferenzangaben Tempmeko 2007, Lake Louise, 21-25, May, 2007, Canada
Freie Schlagworte copper freezing point ; intercomparison ; platinum/palladium thermocouple
Zusammenfassung The objective of the EUROMET project No. 844 in the field of thermometry was the intercomparison of the freezing temperatures of the copper fixed-point cells (t90 = 1084.62 °C) of the participating laboratories by using Pt/Pd thermocouples. For most of the 13 participating laboratories an agreement of the freezing temperatures of the different copper fixed points within ± 0.06 K was found. Furthermore, the results of the intercomparison show that Pt/Pd thermocouples are suitable for use as transfer standards for the dissemination of temperatures and the approximation of the ITS-90 at least up to temperatures of the freezing point of copper.

Zitierung

Edler, F., Anagnostou, M., Bojkovski, J., Gaita, S., Garcia, C., Grudniewicz, E., Helgesen, F., Ivarsson, J., Pauza, A., Rosenkranz, P., Smid, M., Weckström, T., & Zvizdic, D. (2008). Intercomparison of copper fixed-point cells by using Pt/Pd thermocouples [poster]. International Journal of Thermophysics, 29(1), 171–180. https://doi.org/10.1007/s10765-007-0282-z

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag