Zugriffsnummer 20055
Dokumenttyp Zeitschriftenartikel
Sprache Englisch
Titel Characterization of the measurement uncertainty of a laboratory EUV reflectometer for large optics
Autor(in); Institution
Scholze, Frank; 7.1, Photonenradiometrie, PTB-Berlin
Böttger, Th.; IWS Dresden, Fraunhofer-Institut für Werkstoff- und Strahltechnik, Dresden, GERMANY
Enkisch, H.; Carl Zeiss SMT AG, Oberkochen, GERMANY
Laubis, Christian; 7.1, Photonenradiometrie, PTB-Berlin
Loyen van, L.; Europäische Forschungsgesellschaft Dünne Schichten e.V., Dresden, GERMANY
Macco, F.; Carl Zeiss SMTAG, Oberkochen, GERMANY
Schädlich, St.; IWS Dresden, Fraunhofer-Institut für Werkstoff- und Strahltechnik, Dresden, GERMANY
Quelle/Jahr Measurement Science and Technology: 18 (2007), 1, 126 - 130
ISSN 0957-0233
DOI
Verlag Bristol: IOP Publ.
Freie Schlagworte extreme ultraviolet ; metrology ; lithography ; at-wavelength characterization ; reflectometry
Zusammenfassung Quality assurance for the production of optical components for extreme ultraviolet lithography (EUVL) strongly requires at-wavelength metrology. Initially, the at-wavelength characterization of mirrors and masks was developed using synchrotron radiation of electron storage rings, e.g. BESSY II. For the production process of EUV optics, however, the immediate access to metrology tools is necessary and the availability of laboratory devices is mandatory. Recently, an EUV reflectometer (EUVR) for large samples has been put into operation. It consists of a laser-produced plasma (LPP) radiation source, a monochromator and a large goniometer system. It can handle samples with diameters of up to 500 mm, thicknesses of up to 200 mm and weights of up to 30 kg. The wavelength can be varied from 10 nm to 16 nm. The spot size on the sample surface is about 2 mm in both directions. The angle of incidence can be varied from 3° to grazing incidence. In this paper, we describe the measurement uncertainties achieved in the EUVR with reference to measurements by the Physikalisch-Technische Bundesanstalt at its soft x-ray radiometry beamline at BESSY II. Using appropriate reference points for the wavelength calibration of the monochromator, the absolute wavelength uncertainty is below 4 pm in the spectral range from 11 nm to 15 nm. The measured peak reflectance of multilayer mirrors shows a constant offset of about 1% absolute and is reproduced within 0.2%.

Zitierung

Scholze, F., Böttger, T., Enkisch, H., Laubis, C., Loyen van, L., Macco, F., & Schädlich, S. (2007). Characterization of the measurement uncertainty of a laboratory EUV reflectometer for large optics. Measurement Science and Technology, 18(1), 126–130. https://doi.org/10.1088/0957-0233/18/1/015

Exportieren

PTB-Publica Menü

Sprache wechseln: uk flag