| Zugriffsnummer | 19595 |
| Dokumenttyp | Zeitschriftenartikel |
| Sprache | Englisch |
| Titel | CCRI supplementary comparison of standards for absorbed dose to water in Co-60 gamma radiation at radiation processing dose levels |
| Autor(in); Institution |
Burns, D. T.; Bureau International des Poids et Mesures (BIPM), Sèvres, FRANCE
Allisy-Roberts, P. J.; Bureau International des Poids et Mesures (BIPM), Sèvres, FRANCE
Desrosiers, M. F.; National Institute of Standards and Technology, Gaithersburg, MD, USA
Nagy, V. Yu.; National Institute of Standards and Technology, Gaithersburg, MD, USA
Sharpe, P. H. G.; National Physical Laboratory (NPL), Teddington, UK
Laitano, R. F.; Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, Rome, ITALY
Mehta, K.; International Atomic Energy Agency (IAEA), Vienna, AUSTRIA
Schneider, Manfred K.-H.; 6.2, Dosimetrie für die Strahlentherapie, PTB-Braunschweig; PTB-Braunschweig, retired staff
Zhang, Y.; National Institute of Metrology (NIM), Beijing, CHINA
|
| Quelle/Jahr | Radiation Physics and Chemistry: 75 (2006), 9, 1087 - 1092 |
| ISSN | 0969-806X |
| DOI | |
| Verlag | Elsevier Ltd. |
| Zusammenfassung | Six national standards for absorbed dose to water in Co-60 gamma radiation at the dose levels used in radiation processing have been compared over the range from 5 to 30 kGy using the alanine dosimeters of the NIST and the NPL as the transfer dosimeters. The standards are in agreement at the level of around 0.5%, which is significantly smaller than the stated standard uncertainties. |
Zitierung
Burns, D. T., Allisy-Roberts, P. J., Desrosiers, M. F., Nagy, V. Y., Sharpe, P. H. G., Laitano, R. F., Mehta, K., Schneider, M. K.-H., & Zhang, Y. (2006). CCRI supplementary comparison of standards for absorbed dose to water in Co-60 gamma radiation at radiation processing dose levels. Radiation Physics and Chemistry, 75(9), 1087–1092. https://doi.org/10.1016/j.radphyschem.2006.02.003